Description
Product Features:
-
-
- Temperature measure range: -50.0 to 270℃; -58.0 to 518℉
- Resolution: 0.1℃(-50℃~200℃) ; 1℃(200℃~270℃)
- Accuracy: -50℃~ -20℃: ±(reading×1.5%+1℃); -20℃~200℃: ±(reading×1.0%+1℃); 200℃~270℃: ±(reading×2.0%+4℃)
- Response time: T60=5s; T90=10s; T95=12s;
- Measurement Period: 1 time/2 secs
- Display: 1/3 duty, 1/2 bias 3×13 LCD
- Auto power off time: 8 min
- Low Battery Indication: ‘
’ displayed
- Current: ≤1μA(sleep mode) ≤60μA(normal operation)
- Operating temperature: 0~50 ℃, <80% RH
- Storage temperature: -10~60 ℃, <70% RH
- Battery: 1.5V×2
- Dimensions: 181(L)35(W)20(H) mm
- Weight: 65g. Approx. (battery included)
-
Product Parameters:
VA6502 | |
Measurement Range | -50℃~270℃ (-58℉~518℉) |
Resolution | 0.1℃(-50℃~200℃) |
1℃(200℃~270℃) | |
Accuracy | -50℃~20℃:±(reading x1.5%+1℃) |
-20℃~200℃:±(reading x1.0%+1℃) | |
200℃~270℃:±(reading xx2.0%+4℃) | |
Features | |
Auto power off | about 8 minutes |
℃/℉display selection | √ |
Low battery indicator | √ |
Size | 181x35x20 mm |
Weight | ~65g (including battery) |
Application and After-Sales Service:
This digital temperature meter is used to accurately measure internal temperature and ambient temperature.
Name of components:
1. Sensor
2. Battery cover
3.“ON/OFF” key
4.“°C/°F” key
5.LCD display
General Maintenance:
CLEANING INSTRUCTIONS
The meter may be wiped down with a wet sponge or cloth using a mild water based detergent or anti-bacterial soap and rinsed under a gentle stream of cold water.
NOTE:
This unit is not designed for complete submersion or washing in automated dishwashers.
BATTERY REPLACEMENT
Use the following procedure:
When the battery voltage drop below proper operation range the symbol will appear on the LCD display and the battery need to be replaced.
∞Press the battery cover and towards arrowhead direction to open the battery cover.
∞Replace the battery with two new 1.5V batteries (AG13)
Reviews